Performance characteristics of the whole-body discovery IQ PET/CT system

The aim of this study was to assess the physical performance of a new PET/CT system, the Discovery IQ with 5-ring detector blocks. Methods: Performance was measured using the National Electrical Manufacturers Association NU2-2012 methodology. Image quality was extended by accounting for different ac...

ver descrição completa

Detalhes bibliográficos
Autores: Reynés-Llompart, Gabriel; Gámez, Cristina; Romero-Zayas, Immaculada; Rodríguez Bel, Laura; Vercher Conejero, José Luís; Martí-Climent, Josep M.
Tipo de documento: Artigo
Estado:Published version
Data de publicação:2017
País:España
Recursos:Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya)
Repositório:Recercat. Dipósit de la Recerca de Catalunya
OAI Identifier:oai:recercat.cat:2445/132887
Acesso em linha:https://hdl.handle.net/2445/132887
Nível de acesso:Acesso aberto
Palavra-chave:Tumors
Diagnòstic per la imatge
Tomografia per emissió de positrons
Imatge corporal
Diagnostic imaging
Positron emission tomography
Body image
_version_ 1878730612060192769
author Reynés-Llompart, Gabriel
author2 Gámez, Cristina
Romero-Zayas, Immaculada
Rodríguez Bel, Laura
Vercher Conejero, José Luís
Martí-Climent, Josep M.
author2_role author
author
author
author
author
author_browse Gámez, Cristina
Martí-Climent, Josep M.
Reynés-Llompart, Gabriel
Rodríguez Bel, Laura
Romero-Zayas, Immaculada
Vercher Conejero, José Luís
author_facet Reynés-Llompart, Gabriel
Gámez, Cristina
Romero-Zayas, Immaculada
Rodríguez Bel, Laura
Vercher Conejero, José Luís
Martí-Climent, Josep M.
author_role author
collection Recercat. Dipósit de la Recerca de Catalunya
description The aim of this study was to assess the physical performance of a new PET/CT system, the Discovery IQ with 5-ring detector blocks. Methods: Performance was measured using the National Electrical Manufacturers Association NU2-2012 methodology. Image quality was extended by accounting for different acquisition parameters (lesion-to-background ratios [8:1, 4:1, and 2:1] and acquisition times) and reconstruction algorithms (VUE-point HD [VPHD], VPHD with point-spread-function modeling [VPHD-S], and Q.Clear). Tomographic reconstruction was also assessed using a Jaszczak phantom. Additionally, 30 patient lesions were analyzed to account for differences in lesion volume and SUV quantification between reconstruction algorithms. Results: Spatial resolution ranged from 4.2 mm at 1 cm to 8.5 mm at 20 cm. Sensitivity measured at the center and at 10 cm was 22.8 and 20.4 kps/kBq, respectively. The noise-equivalent counting rate peak was 124 kcps at 9.1 kBq/cm3 The scatter fraction was 36.2%. The accuracy of correction for count losses and randoms was 3.9%. In the image quality test, contrast recovery for VPHD, VPHD-S, and Q.Clear ranged from 18%, 18%, and 13%, respectively (hot contrast; 10-mm sphere diameter; ratio, 2:1), to 68%, 67%, and 81%, respectively (cold contrast; 37-mm sphere diameter; ratio, 8:1). Background variability ranged from 3.4%, 3.0%, and 2.1%, respectively (ratio, 2:1), to 5.5%, 4.8%, and 3.7%, respectively (ratio, 8:1). On Q.Clear reconstruction, the decrease in the penalty term (β) increased the contrast recovery coefficients and background variability. With the Jaszczak phantom, image quality increased overall when a reconstruction algorithm modeling the point-spread function was used, and use of Q.Clear increased the signal-to-noise ratio. Lesions analyzed using VPHD-S and Q.Clear had an SUVmean of 6.5 ± 3 and 7 ± 3, respectively (P < 0.01), and an SUVmax of 11 ± 4.8 and 12 ± 4, respectively (P < 0.01). No significant difference in mean lesion volume was found between algorithms. Conclusion: Among the various Discovery bismuth germanium oxide-based PET/CT scanners, the IQ with 5-ring detector blocks has the highest overall performance, with improved sensitivity and counting rate performance. Q.Clear reconstruction improves the PET image quality, with higher recovery coefficients and lower background variability.
eu_rights_str_mv openAccess
format article
id ES_1fd4f7356ab02d13c9cf3b15efc19fce
institution Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya)
instname_str Varias* (Consorci de Biblioteques Universitáries de Catalunya, Centre de Serveis Científics i Acadèmics de Catalunya)
network_acronym_str ES
network_name_str España
oai_identifier_str oai:recercat.cat:2445/132887
publishDate 2017
publishDateSort 2017
publisher The Society of Nuclear Medicine and Molecular Imaging
publisherStr The Society of Nuclear Medicine and Molecular Imaging
reponame_str Recercat. Dipósit de la Recerca de Catalunya
spelling Performance characteristics of the whole-body discovery IQ PET/CT system Reynés-Llompart, Gabriel Gámez, Cristina Romero-Zayas, Immaculada Rodríguez Bel, Laura Vercher Conejero, José Luís Martí-Climent, Josep M. Tumors Diagnòstic per la imatge Tomografia per emissió de positrons Imatge corporal Diagnostic imaging Positron emission tomography Body image The aim of this study was to assess the physical performance of a new PET/CT system, the Discovery IQ with 5-ring detector blocks. Methods: Performance was measured using the National Electrical Manufacturers Association NU2-2012 methodology. Image quality was extended by accounting for different acquisition parameters (lesion-to-background ratios [8:1, 4:1, and 2:1] and acquisition times) and reconstruction algorithms (VUE-point HD [VPHD], VPHD with point-spread-function modeling [VPHD-S], and Q.Clear). Tomographic reconstruction was also assessed using a Jaszczak phantom. Additionally, 30 patient lesions were analyzed to account for differences in lesion volume and SUV quantification between reconstruction algorithms. Results: Spatial resolution ranged from 4.2 mm at 1 cm to 8.5 mm at 20 cm. Sensitivity measured at the center and at 10 cm was 22.8 and 20.4 kps/kBq, respectively. The noise-equivalent counting rate peak was 124 kcps at 9.1 kBq/cm3 The scatter fraction was 36.2%. The accuracy of correction for count losses and randoms was 3.9%. In the image quality test, contrast recovery for VPHD, VPHD-S, and Q.Clear ranged from 18%, 18%, and 13%, respectively (hot contrast; 10-mm sphere diameter; ratio, 2:1), to 68%, 67%, and 81%, respectively (cold contrast; 37-mm sphere diameter; ratio, 8:1). Background variability ranged from 3.4%, 3.0%, and 2.1%, respectively (ratio, 2:1), to 5.5%, 4.8%, and 3.7%, respectively (ratio, 8:1). On Q.Clear reconstruction, the decrease in the penalty term (β) increased the contrast recovery coefficients and background variability. With the Jaszczak phantom, image quality increased overall when a reconstruction algorithm modeling the point-spread function was used, and use of Q.Clear increased the signal-to-noise ratio. Lesions analyzed using VPHD-S and Q.Clear had an SUVmean of 6.5 ± 3 and 7 ± 3, respectively (P < 0.01), and an SUVmax of 11 ± 4.8 and 12 ± 4, respectively (P < 0.01). No significant difference in mean lesion volume was found between algorithms. Conclusion: Among the various Discovery bismuth germanium oxide-based PET/CT scanners, the IQ with 5-ring detector blocks has the highest overall performance, with improved sensitivity and counting rate performance. Q.Clear reconstruction improves the PET image quality, with higher recovery coefficients and lower background variability. The Society of Nuclear Medicine and Molecular Imaging https://hdl.handle.net/2445/132887
spellingShingle Performance characteristics of the whole-body discovery IQ PET/CT system
Reynés-Llompart, Gabriel
Tumors
Diagnòstic per la imatge
Tomografia per emissió de positrons
Imatge corporal
Diagnostic imaging
Positron emission tomography
Body image
status_str publishedVersion
title Performance characteristics of the whole-body discovery IQ PET/CT system
title_full Performance characteristics of the whole-body discovery IQ PET/CT system
title_fullStr Performance characteristics of the whole-body discovery IQ PET/CT system
title_full_unstemmed Performance characteristics of the whole-body discovery IQ PET/CT system
title_short Performance characteristics of the whole-body discovery IQ PET/CT system
title_sort Performance characteristics of the whole-body discovery IQ PET/CT system
topic Tumors
Diagnòstic per la imatge
Tomografia per emissió de positrons
Imatge corporal
Diagnostic imaging
Positron emission tomography
Body image
topic_facet Tumors
Diagnòstic per la imatge
Tomografia per emissió de positrons
Imatge corporal
Diagnostic imaging
Positron emission tomography
Body image
url https://hdl.handle.net/2445/132887